Compass
™
Stainer Operator’s Manual page 37
Running a staining program DOES NOT require a password entry.
To deactivate the password protection, enter 0000 as new password.
Event Log
Event Log utility provides the information regarding various events that took place during the last
run. Every station step and time is recorded. The system captures up to 130 steps and then purges the
oldest as newer events are added.
Press 1 to view the event log. Use the
↑↓
arrows to scroll through the event log. Press STOP to exit.
1. VIEW EVENT LOG
MAKE A SELECTION.
USE ↑↓: TO SCROLL
THROUGH EVENT LOG.
PRESS STOP TO EXIT.
R1 S22 03:00 01624.3
R1 S23 00:00 01810.2
R1 ENDED P01 01812.8
R2 S01 01:00 00020.4
R2 S02 01:00 00086.3
R2 PAUSED 00152.2
R2 RESUMED 00153.3
R2 S03 00:10 00166.0
Rack # Station #
Station Time
(time spent at
the station)
Program Time in seconds
(total accumulated run
time for the rack)
This example is of a run
with multiple racks, using
Program 1
Staining rack 1 finished,
running Program 1.
Next staining rack started.
Run paused and resumed