page 44 Compass
Stainer Operator’s Manual
TROUBLESHOOTING GUIDE
Symptom Probable Causes Suggested Actions
The unit does not
power up
Unit not connected to the
power outlet
No power at the power
outlet
Power Switch in the OFF
position
Ensure that the power cord is
plugged in at the outlet as well as in
the rear of the unit.
Check the circuit breaker for the
power outlet and ensure that the
power is available at the power
outlet.
Turn ON the power switch in the
rear of the.
Blown Fuse Check the fuses and replace any
blown fuse with a fuse of the same
rating. Investigate the cause for
overload and correct the problem
before re-starting the system.
Display does not power
up
Communication between
host and the Display
controller not working or
Display power is
interrupted
Contact Hologic Technical Support
Up-down mechanism
does not initialize or
move
Obstruction in the transport
mechanism, mechanism
does not move freely
Remove any obstructions
Contact Hologic Technical Support
Defective PCB Contact Hologic Technical Support
Defective Motor Contact Hologic Technical Support
Broken Belt Contact Hologic Technical Support
Rotary mechanism does
not initialize or move
Follow the same procedure
as described above for up-
down mechanism
Follow the same procedure as
described above for up-down
mechanism