Normal mA
This menu option returns the system to its normal analogue output state, irrespective of its
previous state. If successful, the following message is displayed for three seconds:
4-20mA Released
Show Status
Thismenuoptiondisplaysthecalibrationandcongurationstatusofthesystem.Thestatus
information is displayed in the following format:
cccccccccccccccc
ssssssss
where: cccccccc... Titleofthecalibrationstatuseld.
sssssssss Current state of calibration.
The different settings from the list are selected using the
and keys on the keypad.
Thecalibrationstatuseldsandvaluesaregivenbelow:
Calibration Status Fields Display Text False State True State
Temperature Cal. Temp Sensor Uncalibrated Calibrated
Zero Calibration Zero Ratio Uncalibrated Calibrated
Span Calibration Span Sensitivity Uncalibrated Calibrated
4-20mA Calibration Analogue Output Uncalibrated Calibrated
Det. Temp. Comp. T Compensation Uncalibrated Calibrated
Det. Sens. Comp. Det Sensitivity Uncalibrated Calibrated
Installation Installation Required Complete
Valid Date Time Date and Time Estimated Set
Loop Test Analogue Check Failed Passed
A.5.4 Diagnose Menu
This sub menu has the following options:
Active Faults Display the active faults in the system.
Active Warns Display the active warnings in the system.
Self Test Perform a diagnostic system self test.
Soft Reset Reset the system.
Active Faults
This menu option displays operational faults when the system is in service.
If there are one or more faults present, then the faults are displayed one at a time.
Use the
and keys to navigate the list.