55
Code
Display Action
Choices
L2.
Hi
(Lo)
Select Limit 2, see Fig. below.
Select with key,
proceed with ENTER
Limit 2 contact response
N/C: normally closed contact
N/O: normally open contact
Select with key,
proceed with ENTER
N/O
(N/C)
Limit 2 switching point
Select with key, edit number with
key, proceed with ENTER
14.00 pH
(
XX.XX pH
XXXX mV)
Limit 2 hysteresis
Select with key, edit number with
key, proceed with ENTER
00.50 pH
(
XX.XX pH
XXXX mV)
Limit 2 delay
The contact is activated with
delay (deactivated without delay)
Select with key, edit number with
key, proceed with ENTER
0010 SEC
(
0 to
9999 SEC)
Limit Lo
Hysteresis +
Switching
point
Contact
0
1
Signal
Limit Hi
Hysteresis -
Signal
Switching
point
Contact
0
1