Mode codes
The mode codes allow fast access to the functions.
Calibration
TA-194.130-HWE01 110903
Key + Code
Description
Cal Info
Display of asymmetry potential and slope
0000
Zero adjustment
of a nominal zero point
1001
Product calibration
Adjustment of asymmetry potential (product)
1105
Adjusting temp probe
1015
Configuration
Error Info Display last error and erase
0000
Configuring
Sensor monitor
Display of uncorrected electrode potential (mV)
Parameter set 1/2 Selecting parameter set
1200
2222
7654
Key + Code
Description
Calibration
Adjustment of asymmetry potential and slope (electrode)
1100
5555
5556
5557
5559
Current source 1 Output current 1 specified
Current source 2 Output current 2 specified
Relay test Manual test of contacts
Manual controller
Manual specification of controller output