GasAlertMicro 5/PID/IR
User Manual
4
Calibrate only in a safe area that is free of hazardous gas, in
an atmosphere of 20.9% oxygen.
Use only BW approved batteries for the GasAlertMicro 5/
PID/IR detectors. Refer to Replacement Parts and
Accessories.
Charge the detector before first-time use. BW recommends
the detector be charged after every workday.
Charge the battery pack immediately when a low battery
alarm occurs.
Read and adhere to the battery cautions provided in Battery
Cautions.
Charge the GasAlertMicro 5/PID/IR batteries using the
recommended charging adapter only. Do not use any other
charging adapter. Failure to adhere to this caution can lead
to fire and/or explosion.
Read and adhere to all instructions in the charger user
manual. Failure to do so can result in fire, electrical shock,
personal injury, and/or property damage.
Extended exposure of the GasAlertMicro 5/PID/IR detectors
to certain concentrations of combustible gases and air may
stress a detector element, which can seriously affect its
performance. If an alarm occurs due to high concentration
of combustible gases, recalibration should be performed, or
if needed, the sensor replaced.
aCautions
Protect the PID sensor from exposure to silicone vapors.
When calibrating O
3
and ClO
2
sensors that are located in
the Toxic 2 sensor position, a single gas calibration cap
must be used to ensure accurate calibration. For more
information, refer to Single Gas Calibration Cap
.
Replace the CO
2
sensor only in a safe and non-hazardous
area that is free of hazardous gas.
The optional pump (M5-PUMP) is certified for use with the
GasAlertMicro 5/PID/IR detectors only.
Do not immerse the detector in liquids.
Do not test the combustible sensor’s response with a
butane cigarette lighter; doing so will damage the sensor.
Do not expose the detector to electrical shock or severe
continuous mechanical shock.
Do not attempt to disassemble, adjust, or service the
detector unless instructions for that procedure are provided
in the user manual and/or that part is listed as a
replacement part. Use only BW Technologies by Honeywell
Replacement Parts and Accessories
.
The detector warranty is void if customer, personnel, or
third parties damage the detector during repair attempts.
Repair attempts made by non-BW Technologies by
Honeywell repair/service personnel voids this warranty.
Lithium Battery Packs
Warning: The lithium battery (M5-BAT08) may present a risk
of fire or chemical burn hazard if misused. Do not
disassemble,
heat above 100°C (212°F), or incinerate.