11. APPENDIX A
104
11.2 FAULT CODES
Number Message Action or Reason
1 Clear log to reset. Log memory error. Clear the event log
2 Clear log to reset. Log memory error. Clear the gas log
3 See manual Memory fault
4 Insert valid cartridge Insert a valid cartridge. If one is fitted,
remove and refit.
5 Insert valid cartridge Memory error. Replace cartridge.
6 Insert valid cartridge An Impact instrument does not
support serviceable cartridges.
7 Insert valid cartridge Sensor mix incorrect. Replace
cartridge for the required type. This is
only generated if the user declines to
accept the cartridge's alarm settings.
12 Check batteries Mixed battery types, i.e. one dry cell,
one rechargeable. Fit two of the same
type.
13 Battery empty. Check
batteries
Battery too low to operate instrument.
Recharge battery or fit new battery
pack.
15 Switch on/off to reset Memory error.
18 See manual Memory error. Incorrect firmware.
21 Contact service centre Memory error. An instrument
parameter is out of the permitted
range.
22 Contact service centre Memory error. A cartridge parameter
is out of the permitted range.