MAN0597 ISSUE 08 - 02/04
2302M5030
4. OPERATION
46
The instrument reports
'Pump Calibration Fail'
The pump was not
blocked correctly
when required.
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked
correctly when
instructed to.
There is a leakage in
the system.
Check that the
following items are
correctly fitted and/
or are not damaged.
Pump Seal
(behind front
cover)
Sample Tube
Pump Adaptor
A blocked condition
was not detected
within 30 seconds of
the start of the test.
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked as
soon as the 'Please
block pump screen'
is displayed.
The Pump Assembly
is faulty.
Request a new
Pump Assembly or
contact Zellweger
Analytics for
assistance.
PUMP FAULT and
WARNING 16 'Pump
Fault' screens are
displayed following the
Pump Calibration.
The Pump
Calibration has
failed.
Remove the Pump
Adaptor and
investigate the fault
(see The instrument
reports 'Pump
Calibration Fail'
above).
Symptom Possible Cause Remedy