MAN0597 ISSUE 08 - 02/04
2302M5030
4. OPERATION
52
stored, printed and analysed. The data can be exported in a
format suitable for use with major spreadsheet software
packages. The datalog memory can be automatically cleared
after a successful download – note that this will not change any
of the alarm levels, instrument settings, or calibration gas
settings or values of the instrument. An internal battery will
retain the data for 5 years even if the instrument battery is
disconnected or the instrument is switched off. Refer to
Section 4.8.1 on how to install the PC software.
4.8.1 Installing the PC Software
(1) Insert the CD into your CD-ROM drive. If the CD
autoruns press the Esc key to stop.
(2) From the Start menu, choose Run.
(3) In the Run dialog box, type d:\English\ setup.exe, where
d is the letter assigned to your CD-ROM drive.
(4) Click OK, then follow the instructions on your screen.
4.8.2 Event Datalogging
All instruments are supplied complete with event datalogging.
This records the time and date whenever an event occurs.
When the datalogging memory is full then the earliest data is
overwritten. In event mode the datalogging memory can store
at least 500 events. An event can be:
• Instrument switch on
• Instrument switch off
• Peak gas reading whilst switched on
• Any gas alarm (A3, A2, A1, STEL, LTEL)
• Low battery
• Fault