|6
1.4.6 Temperature Measurement
• Highly stable, solid state temperature sensor (Thermistor)
• Maximum error over the full temperature range -40° to 70°C (-40 °F to +158 °F) including
linearity, repeatability and hysteresis:
– Temperature: ±0.1% of reading
• Maximum error at reference conditions:
– Temperature: ±0.1% of reading
1.4.7 Pressure Measurement
Ambient temperature range of Pressure Transducer:
• Maximum error over the full temperature range -40° to 70°C (-40 °F to +158 °F) including
linearity, repeatability and hysteresis:
– Pressure: ±0.1% of full scale
• Maximum error at reference conditions:
– Pressure: ±0.04% of full scale
• Long term drift:
– Pressure: ±0.4% of full scale over 10 years
Maximum pressures per pressure transducer type and range: 6 to 1500psig/30-1500psia
Note: For more specification details, please refer EC350 Datasheet.
Type Pmax Units
Gauge 6 psig
Gauge 30 psig
Gauge 60 psig
Gauge 100 psig
Gauge 300 psig
Gauge 600 psig
Gauge 1000 psig
Gauge 1500 psig
Absolute 30 psia
Absolute 100 psia
Absolute 300 psia
Absolute 600 psia
Absolute 1000 psia
Absolute 1500 psia
Pressure Expansion (PE) Board for Pressure Transducer, P3:
1 About MIWI350