RAEGuard 2 PID User’s Guide
22
If there is an error, the “Fault” LED blinks and an error message
blinks. Each alarm condition has a corresponding LED that blinks an
amber color when the readings are outside a specified range or limit.
7.6 Instrument Status Display
After warm-up, the detector enters the Status Display. It automatically
starts testing for errors and goes through a cycle of checking each
alarm condition. If there is no error message or alarm condition, gas
concentration is displayed and LED flashes green.
If there is an error or alarm, a corresponding message shows on the
screen, and the LED indicates different statuses:
Status
External
Alarm
LED LCD
Output
Current
Normal
-
Green
Reading
Reading
Low alarm
Alarm Limit
LOW
Red flashing, 2
Hz
Reading Reading
High alarm
Alarm Limit
HIGH
Red flashing, 3
Hz
Reading Reading
Over Range HIGH
Red flashing, 3
Hz
22mA
Pump Error ERR
Yellow flashing,
1 Hz
2mA
Lamp Error ERR
Yellow flashing,
1 Hz
2mA
Sensor
Negative
Drift
--
Yellow flashing,
1 Hz
2mA
Calibration
Fault
ERR
Yellow flashing,
1 Hz
2mA
Humidity
Sensor Fault
ERR Green, 1 Hz
Reading