Page 58 ST 700 SmartLine Pressure Transmitters User’s Manual Revision 13
Barometric Correct
[27BARCOR]
One point field offset
Calibration Method.
Available for AP models only.
Executing this selection corrects
input pressure to the Barometric
Standard value when the
pressure input is open to the
atmosphere.
The live value of the pressure
input is shown on this display so
the user can easily see the result
of the correction.
Press to
enter menu
selection
Press to
initiate action
Calibrate High Point
[28CAL HI]
The upper pressure value for a
two-point field calibration within
the LRV to URV span.
Cal high point value is in inH2O
[4°C].
Range: LRV to URV
Default: URV
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right
Calibrate Low Point
[29CAL LO]
The lower pressure value for a
two-point field calibration within
the LRV to URV span. Cal low
point value is in inH2O [4°C].
Range: LRV to URV
Default: LRV
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right
Calibrate Adjust Ratio
[30CALRTO]
A ratio value that adjusts the
calibrated curve between the Cal
Low and High Points.
Error Adjustment Ratio
Range: 0.5 to 1.5
Default: 1.0
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right