Page 140 SMV800 Modbus User's Guide Revision 1.0
Calibration
Date
125
DP, SP, PT Calibration Date
in DDMMYY format.
YY : Year with offset as 1900.
Eg. 0x020376 corresponds to
March 2, 2018.
When read, this will always
give Jan 01 2011 as date.
To read Calibration date
configured for DP, SP and PT
refer respective registers.
RW
uint8
Reserve
128
Reserve Address
NA
NA
Model Key
131
Model Key of transmitter
representing the
Differential and Static
Pressure measurement
ranges.
RO
ASCII
Model
Selection
Guide Part 1
135
Model Selection Guide
(MSG) Part 1
RO
ASCII
Model
Selection
Guide Part 2
145
Model Selection Guide
(MSG) Part 2
RO
ASCII
Model
Selection
Guide Part 3
155
Model Selection Guide
(MSG) Part 3
RO
ASCII
Model
Selection
Guide Part 4
165
Model Selection Guide
(MSG) Part 4
RO
ASCII
Comm Module
Temp Up
Count
175
The total number of minutes
that the Communications
Module/board Electronics
Temperature (ET) has
exceeded the upper stress
limit (ET Upper Limit)
RO
uint32
Comm Module
Temp Down
Count
177
The total number of minutes
that the Communications
Modlue/board Electronics
Temperature (ET) has been
below the lower stress limit
(ET Lower Limit)
RO
uint32
Reserve
179
Reserve Address
NA
NA
User Defined
Variables
Monitoring
188
User defined variables limit
check configuration for each
variable.
RW
uint8
Bit - 0 Enable Monitoring
of User Variable 1
Bit - 1 Enable Monitoring
of User Variable 2
Bit - 2 Enable Monitoring
of User Variable 3
Bit - 3 Enable Monitoring
of User Variable 4
Bit - 4 Enable Monitoring
of User Variable 5
Bit - 5 Enable Monitoring
of User Variable 6
Bit - 6 Enable Monitoring
of User Variable 7
Bit - 7 Enable Monitoring
of User Variable 8