Page 164 ST 700 Series HART/DE Option User’s Manual Revision 4.0
Non critical status
(Warning)
(continued)
Non critical status
(Warning)
(continued)
No temperature compensation data exists for
analog output calculations.
Loop accuracy may be slightly compromised.
The effect will be a minor degradation of
ambient temperature influence specifications.
Resolution:
Replace Electronics Module (PWA) to
achieve the maximum current loop accuracy
or return the device to factory for DAC
compensation.
Possible causes:
Factory Calibration is missing. Accuracy will
be compromised.
Resolution:
Replace the Meter Body or return the device
for Factory Calibration.
Sensor Overload/Sensor Fault
Redundant Characterization
Calculation Error
Calculated pressure is above
Upper Transducer Limit (UTL).
Resolution:
Check range and, if required, replace
transmitter with one that has a wider
range. Meter Body may have been
damaged. Check the transmitter for
accuracy and linearity. Replace Meter
Body and recalibrate if needed.
Output current is fixed and not varying with
applied input. Loop current mode is disabled
or Loop Test is active.
Resolution:
Enable Loop Current Mode if it is disabled or
exit the Loop Test mode if active.
Possible causes:
Meter Body temperature is too high.
Accuracy and life span may decrease if it
remains high.
Resolution:
Verify the environment temperature is within
specification. Take steps to insulate Meter
Body from temperature source.
Meter Body Excess
Correct
ZERO or SPAN correction factor is outside
acceptable limits. INPUT may have been
incorrect during calibration or transmitter was