6 - 6
Step 1. Press the Zero key on the remote display several times until Grp 2 Step1 is displayed.
Step 2. Either display will initially display x = NO. Press the Tare key to change this to YES:
Step 3. Press the Zero key to select YES.
Step 4. Press the Tare key until bLd 2 displays:
Step 5. This is the weight range selection against which the load cell is measured. Using the table below, press the Tare
key to rotate around the bLd 2 selections. Press the Zero key to select the bld range desired.
* Common Ranges used as default for most country settings
Step 6. Press the Zero key to display the Cal prompt.
Step 7. Press the Tare key to select YES:
Step 8. Press the Zero key to start the scale calibration.
Step 9. The display shows - - - - - -:
This indicates that the scale is ready to take a zero reading. Check the platter to ensure that it is seated properly
and nothing is touching or interfering with its freedom to move.
Group 2 Step 1 bld Weight Unit Ranges
Bld 0 15 kg Single Range
Bld 1 15 Kg Dual Range*
Bld 2 30 lb Single Range*
Bld 3 30 lb Dual Range
Grp 2
Step 1
21
21 YES
bLd 2
CAL YES
------

Table of Contents