Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
27
104 RPRVOL Reagent prime setup 0
105 CUPVOL Sample cup volume in uL 700
106 WSHTUB Number of execution tube washes 10
107 CUPTYP Not used 0
108 MCYCLE Machine cycle in sec. 12
109 SCHMOD Schedule mode 0
110 WSHSMP Sampling probe wash after every sampling 0 0 = NO, 1 = YES
111 TIMOUT Time out for data communication msec 1500
112 ERTOUT Wait time on error during execution msec 6000
113 BUZMOD Enable buzzer during editing 0
114 LEVREA Else steps before needle stop in the reagent
bottle
7
115 LEVMOD Reagent level check 2 0 = Disable, 1 = Normal,
2 = Normal + Extrapolation
116 LEVSMP Not used 0
117 KLEVR6 Conv. Factor: uL of reagent for sample
needle steps ( rack 6 )
167
118 LEVERR Reagent level error in uL 2000
119 KLEVR9 Conv. Factor: uL of reagent for sample
needle steps ( rack 9 )
113
120 FILTR1 Filter wheel position n°1 340
121 FILTR2 Filter wheel position n°2 0 Double cuvette = 340 nm
122 FILTR3 Filter wheel position n°3 405
123 FILTR4 Filter wheel position n°4 492 Double cuvette = 405 nm
124 FILTR5 Filter wheel position n°5 505 Double cuvette = 492 nm
125 FILTR6 Filter wheel position n°6 546 Double cuvette = 505 nm
126 FILTR7 Filter wheel position n°7 578 Double cuvette = 546 nm
127 FILTR8 Filter wheel position n°8 630 Double cuvette = 578 nm
128 FILTR9 Filter wheel position n°9 700 Double cuvette = 630 nm
140 METTYP Methods structure type 3
159 CODCHK Parameters checksum XXXX