Simple measurements using an HP supplied direct connecting test
xture
In this case, LOAD correction is not required, OPEN/SHORT
correction is enough to correct the residual errors.
Measurements using HP test leads and a test xture.
In this case, CABLE LENGTH and OPEN/SHORT correction
is used. Of course CABLE CORRECTION must be performed
completely describ ed in \Cable Length Selection" in Chapter 4.
Measurement Procedure and Examples 6-17