BIAS Subsystem
:CURRent
The
:CURRent
command sets the DC bias current. Setting
BIAS:CURRent
does not implicitly turn the DC bias ON. The
:CURRent?
query returns the current DC bias current setting value.
This command is eective when the Option 001 DC Bias is installed
or the HP 42841A and the HP 42842A/B are combined.
Command Syntax
BIAS:CURRent
2
:LEVel
3
8
<
:
<
value
>
MIN
MAX
9
=
;
Where,
<
value
>
NR1, NR2, or NR3 format
MIN
Sets or returns the minimum DC bias current level (
0V
)
MAX
Sets or returns the maximum DC bias current level
(When Hi-PW mo de is ON: 100mA)
Note
A sux multiplier and a sux unit, A (amp ere), can be used with
this command. If this query is received when the DC bias level is set
to a current level, error
0
230
Data corrupt or stale
will o ccur.
Example
OUTPUT 717;"BIAS:CURR 10MA"
Query Syntax
BIAS:CURRent
2
:LEVel
3
?
MIN
MAX
Query Response
Returned format is :
<NR3><NL^END>
Example
10 OUTPUT 717;"BIAS:CURR?"
20 ENTER 717;A
30 PRINT A
40 END
8-22 Command Reference