LIST Subsystem
:VOLTage
The
:VOLTage
command clears the previous list sweep p oint table,
and sets the oscillator voltage level sweep points. The
:VOLTage?
query returns the current settings of the voltage sweep p oints.
Command Syntax
LIST:VOLTage
<
value
>
2
,
<
value
>
*
3
*Max. 10 sweep p oints
Where,
<
value
>
is the NR1, NR2, or NR3 format
Example
OUTPUT 717;"LIST:VOLT 1.5" !Set 1.5V to point 1
OUTPUT 717;"LIST:VOLT 1E-2,2E-2,3E-2,4E-2"
! Set 10 mV to point 1, .... 40 mV to point 4
Note
A sux multiplier and a sux unit,
V
(voltage), can be used with
this command.
Query Syntax
LIST:VOLTage?
Query Response
Returned format is :
<NR3>
2
,<NR3>
*
3
<NL^END>
Example
10 DIM A$[100]
20 OUTPUT 717;"LIST:VOLT?"
30 ENTER 717;A$
40 PRINT A$
50 END
Note
If this query is received when the List Sweep parameter is set to
anything other than voltage, error
0
230
Data corrupt or stale
will
o ccur.
Command Reference 8-33