STATus Subsystem
:OPERation:ENABle
The
:OPERation:ENABle
command sets the enable bits of the
standard op eration status event register that allows true conditions
in the event register to b e rep orted in the summary bit of the status
byte register. The
:OPERation:ENABle?
query returns the current
setting of the enable bits of the op eration status event register.
Command Syntax
STATus:OPERation:ENABle
<
value
>
Where,
<
value
>
NR1 format : decimal expression of enable bits of the
operation status event register
The denition of each bit in the op eration status event register is as
follows.
Bit No. Description
15 { 5 Always 0 (zero)
4 Measurement Complete Bit
3 List Sweep Measurement Complete Bit
2, 1 Always 0 (zero)
0 Correction Data Measurement Complete Bit
Example
OUTPUT 717;"STAT:OPER:ENAB 16"! Bit 4 enable
Query Syntax
STATus:OPERation:ENABle?
Query Response
Returned Format is:
<
value
>
<NL^END>
Example
10 OUTPUT 717;"STAT:OPER:ENAB?"
20 ENTER 717;A
30 PRINT A
40 END
Command Reference 8-81