30. Connect the 2 m Test Leads (HP 16048D) to the
UNKNOWN
terminals.
31. Press the
4
MEAS SETUP
5
MENU key and the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
softkey
to display the
CORRECTION
page.
32. Set the
CABLE
to 2 m.
33. Store the
OPEN CORRECTION
data and the
SHORT
CORRECTION
data referring to Step 4 through 9. In this
pro cedure the OPEN termination and the SHORT termination
should b e connected to the 2 m Test Leads (HP 16048D).
34. Set the HP 4284A to the Manual Trigger mo de.
35. Set the measurement function to
Cp-D
.
36. Connect the 1000 pF standard to the 2 m Test Leads (HP
16048D).
37. Perform Steps 38 through 39 for all the test frequencies listed in
Table 10-10.
38. Press the
4
TRIGGER
5
key.
39. Conrm that the HP 4284A's reading is within the test limits in
Table 10-10.
40. Connect the 4 m Test Leads (HP 16048E) to the
UNKNOWN
terminals.
41. Press the
4
MEAS SETUP
5
MENU key and the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
softkey
to display the
CORRECTION
page.
42. Set the
CABLE
to 4 m.
43. Store the
OPEN CORRECTION
data and the
SHORT
CORRECTION
data referring to steps 4 through 9. In this
pro cedure the OPEN termination and the SHORT termination
should b e connected to the 4 m Test Leads (HP 16048E).
44. Set the HP 4284A to the Manual Trigger mo de.
45. Set the measurement function to
Cp-D
.
46. Connect the 1000 pF standard to the 4 m T
est Leads (HP
16048E).
47. Perform Steps 48 through 49 for all the test frequencies listed in
Table 10-10.
48. Press the
4
TRIGGER
5
key.
49. Conrm that the HP 4284A's reading is within the test limits
listed in Table 10-10.
10-18 Performance Tests