Multimeter Reading
(Signal Frequency: 960 kHz, Hi PW: OFF)
PASS [ ] FAIL [ ]
OSC
LEVEL
MULTIMETER
READING
MINIMUM
(0.97
2
M.R.
0
0.5 mV)
ACTUAL MAXIMUM
(1.03
2
M.R.+0.5 mV)
5mV
10 mV
20 mV
50 mV
100 mV
200 mV
250 mV
500 mV
1V
2V
M.R. is the Multimeter Reading for the HP 4284A's test signal lev
el.
Multimeter Reading
(Signal Frequency: 1.25 kHz, Hi PW: ON)
Option 001 ONLY
PASS [ ] FAIL [ ]
Level Monitor Reading
(Signal Frequency: 1.25 kHz, Hi-PW: ON)
Option 001 ONLY
OSC
LEVEL
MULTIMETER
READING
MINIMUM
(0.97
2
M.R.
0
0.5 mV)
ACTUAL MAXIMUM
(1.03
2
M.R.+0.5 mV)
(0.97
2
M.R.
0
0.5 mV) (1.03
2
M.R.+0.5 mV)
10 mV
(0.97
2
M.R.
0
0.5 mV) (1.03
2
M .R.+0.5 mV)
1V
(0.97
2
M.R.
0
5 mV) (1.03
2
M.R .+5 mV)
2.5 V
(0.97
2
M.R.
0
5 mV) (1.03
2
M. R.+5 mV)
20 V
10-38 Performance Tests