Impedance
Measurement Accuracy
Test
10 pF Standard
C.V.: (Cal. Value at 1 kHz)
OSC Level: 510 mV
SIGNAL
FREQUENCY
MINIMUM ACTUAL MAXIMUM
12.5 kHz Cp C.V.
0
0.0318 pF C.V. +0.0318 pF
D
0
0.0031 0.0031
48 kHz Cp C.V.
0
0.0246 pF C.V. +0.0246 pF
D
0
0.0028 0.0028
96 kHz Cp C.V.
0
0.0275 pF C.V. +0.0275 pF
D
0
0.0035 0.0035
1 MHz Cp C.V.
0
0.0249 pF
C.V. +0.0249 pF
D
0
0.0038 0.0038
OSC Level: 20 mV
PASS [ ] FAIL [ ]
Performance Tests 10-41