Chapter 4 Command Reference
Introduction
Programming Guide 4-69
Descriptions of the Measurement Functions<function> (Cont.)
:MEASure[:SCALar][:VOLTage]:PERiod? (Cont.)
Ch3<expected_value>
range: 0.33 ns to 10.0 ns
resolution: <expected_value> should be within 10% of input period
for optimum arming configuration
default: 2 ns
Ch3 <resolution>
description: value indicates decade corresponding to least significant
digit of the result
range: 1E-24 to 1E-11 sec
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
resolution: <resolution> should use a mantissa of 1.0 and be an even
power of 10
default: value which indicates 4 digits of resolution for the
specified <expected_value>
<source_list>
range: (@1) | (@2) | (@3)
default: (@1)