Chapter 4 Command Reference
Introduction
Programming Guide 4-87
[:SENSe]:FREQuency:ARM:STOP:SOURce IMMediate | EXTernal
| TIMer | DIGits
Sets or queries the stop arm for Frequency, Frequency Ratio, and Period
measurements.
A sequence of ASCII-encoded bytes: IMM, EXT, TIM, or DIG
*RST: TIMer
Gate & ExtArm
[:SENSe]:FREQuency:ARM:STOP:TIMer
<numeric_value> [S]
Sets or queries the gate time used in arming Frequency, Frequency Ratio, and
Period measurements.
For short gate time: 1E-3 to 99.99E-3 seconds
For long gate time: 100E-3 to 1000.000 seconds
For short gate time: 0.01E-3 seconds
For long gate time: 1E-3 seconds
Numeric data transferred as ASCII bytes in <NR3> format with six significant
digits.
*RST: 100E-3 S
Only applies when [:SENS]:FREQ:ARM:STOP:SOUR TIM is selected.
Gate & ExtArm
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
<numeric_value
Resolution
Query Response
Comments
Related
Front-Panel
Keys