Chapter 4 Command Reference
[:SENSe] Subsystem
4-102 Programming Guide
[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SLOPe POSitive |
NEGative
Sets or queries the slope of the external stop arm signal used in external arming
time interval (including Time Interval, Risetime, Falltime, Duty Cycle, and Pulse
Width functions) measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: POSitive
Only applies when [:SENS]:TINT:ARM:EST OP:LAY2:SOUR EXT is
selected.
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SOURce IMMediate |
EXTernal
Sets or queries the stop arm for time interval (including Time Interval, Risetime,
Falltime, Duty Cycle, and Pulse Width functions) measurements.
A sequence of ASCII-encoded bytes: IMM or EXT
*RST: IMMediate
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:ECOunt
<numeric_value>
Sets or queries the number of events used to delay the stop arm for Time Interval
measurements.
1 to 99,999,999
1
Numeric data transferred as ASCII bytes in <NR1> format.
*RST: 1
Only applies when [:SENS]:TINT:ARM:ESTOP[:LAYer[1]]:SOUR
INTernal2 is selected.
Query Response
Comments