Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide 2-31
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
<function> *
<parameters>
[,<source_list>] **
Std/
New
[:VOLTage]:DCYCle
or
[:VOLTage]:PDUTycycle
[:VOLTage]:FALL:TIME
or
[:VOLTage]:FTIMe
[:VOLTage]:FREQuency
[:VOLTage]:FREQuency:RATio ***
[<reference>]
[<reference>]
[<lower_reference>[,upper_reference>]
]
[<lower_reference>[,upper_reference>]
]
[<expected_value>[,<resolution>]]
[<expected_value>[,<resolution>]]
[ (@1) ]
[ (@1) ]
[ (@1) ]
[ (@1) ]
[ (@1) | (@2) | (@3)
]
[ (@1), (@2 | @3) |
(@2 | @3), (@1) ]
Std
Std
Std
Std
Std
New
[:VOLTage]:MAXimum [ (@1) | (@2) ]
Std
[:VOLTage]:MINimum [ (@1) | (@2) ]
Std
[:VOLTage]:NWIDth [<reference>] [ (@1) ]
Std
[:VOLTage]:PERiod [<expected_value>[,<resolution>]] [ (@1) | (@2) | (@3)
]
Std
[:VOLTage]:PHASe [ (@1), (@2) ]
Std
[:VOLTage]:PTPeak [ (@1) | (@2) ]
Std
[:VOLTage]:PWIDth [<reference>] [ (@1) ]
Std
[:VOLTage]:RISE:TIME
or
[<lower_reference>[,upper_reference>]
]
[ (@1) ]
Std
[:VOLTage]:RTIMe [<lower_reference>[,upper_reference>]
]
[ (@1) ]
Std
[:VOLTage]:TINTerval [ (@1), (@2) ]
New
[:VOLTage]:TOTalize:CONTinuous ****
[ (@1) ]
New
[:VOLTage]:TOTalize:TIMed [<gate_time>] [ (@1) ]
New
______________________________
* The only functions which can be derived (using FETC? or READ?) from the stored data are period
to/from frequency, maximum to/from minimum, maximum to/from peak-to-peak, and minimum
to/from peak-to-peak. All other functions require an acquisition of a new type.
** <source_list> has the same syntax as SCPI <channel _list> syntax. For example, a
single-channel function (e.g., frequency, period, etc.) would use (@1) to specify channel 1,
whereas a two-channel function (e.g., time interval, phase, and frequency ratio) would use (@1),
(@2) to specify a measurement between channel 1 and channel 2.
***For HP 53131A firmware revision s below 3335, only Ratio 1 to 2 and Ratio 1 to 3 were offered.
**** This <function> is only allowed with :CONFigure.