EasyManua.ls Logo

HP 53131A/132A 225 MHz - Page 65

HP 53131A/132A 225 MHz
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide 2-41
Table 2-3. HP 53131A/132A *RST State (Continued)
Command Header Parameter State
*
EMC
<NRf> 0 (i.e., disabled)
:FORMat[:DATA] ASCii | REAL ASCii
:HCOPy:CONTinuous <Boolean> OFF
:INITiate:AUTO
:INITiate:CONTinuous
<Boolean>
<Boolean>
OFF
OFF
:INPut[1|2]:ATTenuation
:INPut[1|2]:COUPling
:INPut[1|2]:FILTer[:LPASs]:STATe
:INPut[1|2]:IMPedance
1 | 10
AC | DC
<Boolean>
<numeric_value> [OHM]
1
AC
OFF
1E6 OHM
[:SENSe]:EVENt[1|2]:HYSTeresis:RELative
[:SENSe]:EVENt[1|2]:LEVel[:ABSolute]:AUT
O
[:SENSe]:EVENt[1|2]:LEVel:RELative
[:SENSe]:EVENt[1|2]:LEVel:SLOPe
[:SENSe]:EVENt2:FEED
[:SENSe]:FREQuency:ARM[:STARt]:SLOPe
[:SENSe]:FREQuency:ARM[:STARt]:SOURce
[:SENSe]:FREQuency:ARM:STOP:DIGits
[:SENSe]:FREQuency:ARM:STOP:SLOPe
[:SENSe]:FREQuency:ARM:STOP:SOURce
[:SENSe]:FREQuency:ARM:STOP:TIMer
[:SENSe]:FREQuency:EXPected[1|2|3]:AUTO
[:SENSe]:FUNCtion[:ON]
[:SENSe]:PHASe:ARM[:STARt]:SLOPe
[:SENSe]:PHASe:ARM[:STARt]:SOURce
[:SENSe]:ROSCillator:EXTernal:CHECk
[:SENSe]:ROSCillator:SOURce:AUTO
<numeric_value> [PCT]
<Boolean>
<numeric_value> [PCT]
POSitive | NEGative
INPut1 | INPut2
POSitive | NEGative
IMMediate | EXTernal
<numeric_value>
POSitive | NEGative
IMMediate | EXTernal | TIMer | DIGits
<numeric _value> [S]
ON
<sensor_function>
POSitive | NEGative
IMMediate | EXTernal
ON | OFF | ONCE
<Boolean>
100 PCT *
ON
50 PCT
POSitive
INPut2
(i.e., separate mode)
POSitive
IMMediate
4
NEGative
TIMer
100E3 S
ON
FREQuency 1
POSitive
IMMediate
ON
ON
________________________________
*For prior firmware revisions (3317, 3335, and 3402), *RST sets up 100% meaning maximum
sensitivity.
For current firmware revision, *RST sets up 0%meaning minimum
hysteresis.

Table of Contents

Related product manuals