12
Table 2-1. Test Equipment Required
TYPE REQUIRED CHARACTERISTICS USE RECOMMENDED MODEL
Oscilloscope Sensitivity: 1 mV
Bandwidth: 20MHz & 100MHz
Input: differential, 50
Ω
& 10M
Ω
P,T HP54504A
RMS Voltmeter True rms, 10MHz bandwidth
Sensitivity: 1 mV Accuracy: 5%
P HP3400A
Logic Pulser 4.5 to 5.5Vdc @ 35mA T HP546A
Multimeter Resolution: 100nV
Accuracy: 0.0035%, 6½ digit
P,C,T HP 3458A
Signature Analyzer -- T HP 5004A
HP-IB Controller Full HP-IB capabilities C,T,P HP Series, 200/300
Current Probe No saturation at 30Adc
Bandwidth: 20Hz to 20MHz
P Tektronix P6303 Probe/
AM503 Amp/
TM500 Power Module
Electronic Load Voltage range: 60Vdc P,C HP 6060A
Current range: 30Adc
Power range: 250W
Open and short switched
Power Resistor*
Value: 0.25
Ω
>200W (6033A)
P,C
Value: 2.3
Ω
>200W (6038A)
Current-Monitoring
Shunts
Value: 10m
Ω
±
0.04% @ 100W (6033A)
PC: 0.0004%/W
P,C** Guildline 9230/100
Value: 100m
Ω
±
0.04% @ 25W (6038A)
PC: 0.0004%/W
Guildline 9230/15
Calibration and Test
Resistors
Value: 100
Ω
, 5%, 1W
1
Ω
, 5%, ½W
1K
Ω
, 5%, ¼W
2K
Ω
, 0.01%, ¼W
C,T
Terminating
Resistors (2)
Value: 50
Ω
±
5%, noninductive
P
Blocking
Capacitors (2)
Value: 0.01
µ
F, 100Vdc
P
Common-mode
Toroidal Core
≥
3.7
µ
H/turn
2
≅
23mm I.D.
P Ferrox-Cube
500T600-3C8,
HP 9170-0061
Switch* SPST, 30A @60V P
DC Power Supply Voltage range: 0-60Vdc
Current range: 0-3Adc
C,T HP 6024A
P = performance testing C = calibration adjustments T = troubleshooting
* Not required if using electronic load.
** Less accurate, and less expensive, current-monitor resistors can be used, but the accuracy to which current programming
and readback can be checked must be reduced accordingly.