i.
Reading on
differential
voltmeter
should
not
vary
from
reading
recorded
in
step
g
by
more
than
the
following
(variations
expressed
in
mVdc):
Model
6282A
628SA
6286A 6290A 6291A
6296.A
Variation
:2
:±3
:1:3
:5
±5
:7
5-18
Ripple and Noise.
To
check
the
ripple
and
noise,
proceed
as
follows:
a.
Retain
test
setup
used
for
previous
line
regulation
test
except
connect
AC
voltmeter
across
output
terminals
as
shown
in Figure
5-6.
b. Adjust
variable
auto
transformer
for
125
VAC
input.
c.
Set
METER
switch
to
highest
current
range.
d. Turn
CURRENT
controls
fully
clockwise
and
adjust
VOLTAGE
controls
until
front
panel
meter
indicates
exactly
the maximum
rated
output
voltage.
e.
AC
voltmeter
should
read
less
than
O.
50mVrms.
POWER
SUPPLY
UNDER
'ltST
LOAO
RtSISTORS
AC
VOLTMma
,,
403
Figure
S-6. Ripple and
Noise.
Constant
Voltage
5-19
Transient
Recovery Time. To
check
the
transient
recovery
time
proceed
as
follows:
a.
Connect
test
setup
shown
in
Figure
5-7.
b. Turn
CURRENT
controls
fully
clockwise.
c.
Set
METER
switch
to
highest
current
range
and turn on
supply
. d. Adjust
VOLTAGE
~ontrols
until
front
panel
meter
indicates
exactly
the
maximum
rated
output
current
or S
amperes,
whichever
is
smaller.
e.
Close
line
switch
on
repetitive
load
switch
setup.
5-5
POWtR
SUPPLY
UNDER
TEST
+
OSCILLOSCOPE
,,
140A
0
0-+--------------------1-.0
G
0
CONTACT
PROTECTION
Ntl'WORIC
Ry
l..ld
I"
SW
200V
(NOTE
31
Rx
f----·Vlll.--..l\/V\.--1
~ODEL
NO.
RESIS
A~·
•~u~••
Rx
a.,
6282A
0.10
0.90
628SA
0.20
3. 80
6286A
0.10
1.90
6290A
0.
33
13. 0
6291A
0.20
7. 80
6296"
0.
33
I
1'.
66
.,
I
NOTES:
I.
THIS
DRAWING
SHOWS
A SVCGESTED
METHOD
or
BUILDING A
L~D
SWITCH.
HOWMR.
OTHER
METHODS
COULD
IE
USED:
SUCH
AS
A
TRANSISTOR
SWITCHING
NETWORK.
MAXIMUM
LOAD
RAnNGS or
LOAD
SWITCH
ARE;
S AMPS.
SOOV.
ZSOW
INOT
2S0'1Wl
2.
USE
MERCURY
REIAY:
CURE
TYPE
HGP
I
002
OR
W.
E.
TYPE
2768.
l.
USE
Wtat
WOUND
RC.~ISTOR.
Figure
S-7.
Transient
Response,
Test
Setup
f.
Adjust
2SK
potentiometer
until
a
stable
dis
play
is
Qbtained
on
oscilloscope.
Waveform
should
be
within
the
tolerances
shown
in
Figure
S-8
(output
should
return
to
within
15
mV
of
original
value
in
less
than
50
microseconds).
I
1-so..stc-t
UNLOADING
TRANSIENT
f---so
....
~
J
__
I I
ISMV
ov-r-----__,,,c....
____
,i;;:::;::;._-r--
~DING
TRANSIENT
Figure
5-8.
Transient
Response
Waveforms
_:s-20
Output
Impedance.
To
check
the
output
im-
pedance,
proceed
as
follows:
a.
Connect
test
setup
shown
in
Figure
S-9