SCPI
COMMAND SUMMARY
l
DIAGnostics:TEST[:EXECute]
<mm>
The specified
selftest
is executed. Normal instrument operation is
suspended and the instrument state is restored upon exiting the
selftest
mode.
l
DIAGnostics:TEST:LOG:SOURce
ALLIFAIL
l
DIAGnostics:TEST:LOG:SOURce?
Sets and queries the source for the raw data logging. ALL specifies
that all raw data points are displayed. FAIL selects only those data
points out of the test limits. Both commands are executable in
selftest
mode. After
*RST,
the setting is FAIL.
l
DIAGnostics:TEST:LOG[:STATe]?
l
DIAGnostics:TEST:LOG[:STATe]
ONlOFFlllO
Selects and queries the raw data logging ON/OFF switch. Both
commands are executable in
selftest
mode.
After
*RST,
the setting is 0.
l
DIAGnostics:TEST:LOOP ONlOFFlllO
l
DIAGnostics:TEST:LOOP?
Selects and queries the test looping ON/OFF switch. Both
commands are executable in
selftest
mode.
After
*RST,
the setting is 0.
l
DIAGnostics:TEST:NAME?
[<nun01
Queries the name of a
selftest
by number. If the number is not
specified then an array of all the
selftest
names is returned.
l
DIAGnostics:TEST:POINts?
Returns the number of points of
selftest
that is output using
DIAGnostics:TEST:NAME?orDIAGnostics:TEST:RESult?.
l
DIAGnostics:TEST:RESult?
[<num>]
Queries the result of a selftest, by number. The response is a string
containing either
Passed, Failed,
or
NotRun.
If <num> is missing,
an array of
selftest
results are returned.
l
DIAGnostics:TINT?
<mm>
A test feature that returns the value passed to it. This is used to test
the HP-IB interface.
l
DISPlayC:STATel ONlOFFlllO
l
DISPlay
[
: STATeI ?
Sets and queries the display ON/OFF switch.
After
“RST,
the value is 1.
Operating and Programming Reference S-33