These commands output the power meter calibration array. Values
should be entered as 100 times the power meter reading in
dB.
A
default array is used if a power meter calibration sweep, TAKCS, has not
been taken:
OUTPPMCALl
Channel 1.
OUTPPMCALZ
Channel 2.
These 4 commands output the pre-raw measurement data. Analogous
to OUTPRAW except that pre-raw data has not had sampler correction
nor attenuator offsets applied. These offsets are not necessary for data
that will be fully error corrected. See BASIC programming Example 2E:
‘lake4
-
Error Correction Processed on an External Computer. The
arrays hold Sl
1,
S2
1,
S12,
and
S22,
respectively:
OUTPPRE
1
OUTPPRE2
OUTPPRE3
OUTPPRE4
Array 1
(Sll
data).
Array 2
(S21
data).
Array 3
(S12
data).
Array 4
(S22
data).
OUTPPRIN
Outputs a raster dump of the display,
intended for a graphics printer.
OUTPPRNALL
Outputs all of the List Values or Operating
Parameters in text mode. Activate the
desired function by preceding with LISV
or OPEP, respectively.
HP-IB
Progr
amming
and Command Reference Guide
l-137