Error Messages
Fehlermeldungen
Error Message Root Causes Corrections
EE
Measurement Error Exceeding measurement
range.Measure again
quietly. Wrap the cuff
correctly and keep wrist
steady during
measurement. If error
keeps occurring, return the
device to the local
distributor or service
center.
E1
Air Circuit Abnormality Measure again quietly. If
error keeps occurring,
return the device to the
local distributor or service
center.
E2
Pressure Exceeding 300
mmHg
Switch the unit off and
measure again quietly. If
error keeps occurring,
return the device to the
local distributor or service
center.
E3
Data Error Remove and reload the
batteries. If error keeps
occurring, return the
device to the local
distributor or service
center.
Fehlermeldung Ursachen Korrekturen
EE
Messfehler
Überschreiten des
Messbereichs.
Messen Sie erneut in
Ruhe. Legen Sie die
Manschette korrekt an und
halten Sie das Handgelenk
während der Messung still.
Tritt der Fehler erneut auf,
geben Sie das Gerät an
das lokale Vertriebs- oder
Service Center zurück.
E1
Luftkreislaufstörung Messen Sie erneut in
Ruhe. Tritt der Fehler
erneut auf, geben Sie das
Gerät an das lokale
Vertriebs- oder Service
Center zurück.
E2
Der Druck überschreitet
300 mm Hg
Schalten Sie das Gerät
aus und messen Sie
erneut in Ruhe. Tritt der
Fehler erneut auf, geben
Sie das Gerät an das
lokale Vertriebs- oder
Service Center zurück.
E3
Datenfehler Entfernen und setzen Sie
die Batterien wieder ein.
Tritt der Fehler erneut auf,
geben Sie das Gerät an
das lokale Vertriebs- oder
Service Center zurück.