Chapter 1 Introduction
ET-7000 AIO User Manual, Apr. 2009, V1.04, EMH-011-104----------44
It is not recommended that calibration be performed until
the process is fully understood.
1.8.2. The calibration procedure for ET-7017 and ET-7017-10:
1. Warm up the module for at least 30 minutes.
2. Set the type code to the type you want to calibrate.
3. Enable calibration.
4. Apply the zero calibration voltage/current to channel 0.
5. Enable the zero calibration register.
6. Apply the span calibration voltage/current to channels 0.
7. Enable the span calibration register.
8. Disable calibration.
Table 1.8.5 Modbus address for calibration:
Register Points Description Access Type
00830 1
Enable/Disable Calibration
0=Disabled
1=Enabled
R/W
00831 1 Performs a zero calibration, 1 = run. W (Pulse)
00832 1 Performs a span calibration, 1 = run. W (Pulse)
Table 1.8.6 Calibration voltage/current used by the ET-7017 and ET-7017-10 series:
Type code
08
09 0A 0B 0C 0D
Zero Input
0V
0V 0V 0mV 0mV 0mA
Span Input
+10V
+5V +1V +500mV +150mV +20mA