Electrical Specifications
Función Rango Rango Resolución Precisión
61-747 61-757 ±(a%+b)
Voltaje CA (V)
TRMS
4.000V 6.000V 0.001V
±(1.0%+5)40.00V 60.00V 0.01V
400.0V 600.0V 0.1V
600V 1000V 1V ±(1.5%+5)
VCA LoZ* (V) NA 600.0V 0.1V ±(2.5%+5)
Corriente CA (A)
TRMS
40.00A 60.00A 0.01A ±(2.0%+9)
400.0A 600.0A 0.1A ±(2.0%+5)
Frecuenci a (ACA
Hz)
99.99Hz 99.99Hz 0.01Hz
±(0.1%+5)
999.9Hz 999.9Hz 0.1Hz
Voltaje CD (V)
400.0mV 600.0mV 0.1mV ±(1.0%+8)
4.000V 6.000V 0.001V
±(1.0%+5)4.000V 60.00V 0.01V
400.0V 600.0V 0.1V
600V 1000V 1V ±(1.5%+5)
Corriente CD (A)
40.00A 60.00A 0.01A ±(2.0%+9)
400.0A 600.0A 0.1A ±(2.0%+5)
Resistencia (Ω)
400.0Ω 600.0Ω 0.1Ω
±(1.5%+5)
4.000kΩ 6.000kΩ 0.001kΩ
40.00kΩ 60.00kΩ 0.01kΩ
400.0kΩ 600.0kΩ 0.1kΩ
4.000MΩ 6.000MΩ 0.001MΩ
40.00MΩ 60.00MΩ 0.01MΩ ±(2.0%+5)
1. Nota: * denota una función sólo del 61-757.
2. Protección de Sobrecarga: 1000VRMS y 600 ACA
3. La precisión a es un % de la lectura y b es el DMS (dígito menos significativo).
4. Factor de cresta AC 2.5 en conteo de 4000, y onda no sinusoidal según factor
de onda calculado de la siguiente manera:
b) Agregar 3% cuando el factor de cresta sea 1~2
b) Agregar 5% cuando el factor de cresta sea 2~2.5