In
local
operation, the limit power value
is
read
by
selecting the
LIM
P
parameter,
in
the
PARAMETER
area
of the hnt panel,
and
visually
reading
the LASER
display.
EXAMPLES
"LAS:LIM:MDP?" -response: 3.0, means the monitor PD power limit
is
set to 3.0 mW.
":LAS:Limit:MDP?" -response: 10.0, means
the
monitor PD power limit
is
set to 10.0 mW.
Front Panel
Remote
-
The LASer:MODE? query
returns
the selected
laser
control mode.
SYNTAX DIAGRAM
PARAMETERS None. The
response
will
be
in
the form:
-where the response
is
character
data.
POINTS OF
INTEREST
I,
mode
is
the same
as
I mode (low bandwidth), except that the output low
bandpass
filter
is
disabled in
I,
mode.
In
local
mode, the LASER control mode
is
indicated
by
the
LED
in the LASER MODE
area
of the fiont panel.
If
the
P
mode LED
is
lit,
the mode
is
I,,
when the
CAL
PD
value
is
zero,
and
the
mode
is
P,,
when the
CAI.
PD
value
is
non-zero.
EXAMPLES
"LAS:MODE?" -response: Ilbw,
means
&t
constant I (current) mode
is
in
effect for the
laser output.
":Las:ModelN -response: MDP, means
that
constant P (power) mode
is
in
effect
for the laser
output, and
CALMD
>
0.
"LaxMode?"
-response:
MDI,
means
that
constant
P
@ewer)
mode
is
in
effect
for the
laser
output, and
CALMD
=
0.
"Laser:MODE?"
-response:
Ihbw, means that constant I,, (current,
high
bandwidth) mode
is
in
effect
for the laser output.