SYNTAX
DIAGRAM
PARAMETERS
An
<nrf
value>
which represents the photodiode
feedhack
power, in
mW.
POINTS
OF
INTEREST
If
the CALMD (CAL PD) parameter
is
set
to
0,
the
LAS:MDP
value
will
not
be
used.
In
this
case,
the
measured
MDP would
be
invalid,
and
the MDI
(I,)
set
point would
be
used.
In
local
mode,
the photodetector power
set
point
would
be
set
by
first
pressing the (LASER
MODE) P switch to enter constant power mode, and then pressing the
SET
switch,
and
then
turning the
ADJUST
knob
until
the
desired
value
appeared
on the
LASER
display.
EXAMPLES
"LasMdp
40"
-action: The LASER output
is
controlled
so
that the photodiode
feedback
power
remains
constant
at
40
mW.
"Laser:MDP
200"
-action: The LASER output
is
controlled
so
that
the photodiode
feedback
power
remains
constant
at
200
mW.
Front
Panel
The
LASer:MDP?
query returns the
value
of the laser photodetector power measurement, in
mW.
SYNTAX
DIAGRAM
n
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ILX Lightwave LDC-3700 Series and is the answer not in the manual?

ILX Lightwave LDC-3700 Series Specifications

General IconGeneral
BrandILX Lightwave
ModelLDC-3700 Series
CategoryController
LanguageEnglish

Related product manuals