Loretta HensonAug 2, 2025
What to do if my iM3 42-12 has no power?
- AAlexander MillerAug 2, 2025
If your iM3 Media Converter has no power, the cause may be a blown fuse. Replace the fuse with a slow blow 400mA – 250 v fuse.
What to do if my iM3 42-12 has no power?
If your iM3 Media Converter has no power, the cause may be a blown fuse. Replace the fuse with a slow blow 400mA – 250 v fuse.
Why is the scaler tip vibrating but there is no coolant spray in my iM3 42-12?
If the scaler tip on your iM3 Media Converter vibrates but there's no coolant spray, it could be due to a blocked coolant line or a closed flow control valve. Check the filter and open the flow control valve.
How to fix poor or no tip vibration on iM3 42-12 Media Converter?
If you're experiencing poor or no tip vibration with your iM3 Media Converter, it could be due to several reasons: * Water in the handpiece: Empty the handpiece. If the issue reoccurs, replace the handpiece. * Loose Ferrite rod: Tighten the Ferrite Rod. * Broken Ferrite Rod: Replace the Ferrite Rod. * Worn out tip: Replace the tip.
What to do if there is coolant in the interior of my iM3 Media Converter Handpiece?
If you find coolant in the interior of your iM3 Media Converter handpiece, it is likely due to a defective O-ring in the handpiece. Replace the O-ring to resolve this issue.
Why is there a coolant leak between the handpiece and scaler tip insert in my iM3 42-12 Media Converter?
If coolant is leaking between the handpiece and scaler tip insert of your iM3 Media Converter, this is likely caused by a defective O-ring in the handpiece. Replace the O-ring to stop the leak.
How to fix coolant leaking between handpiece and connectors on iM3 42-12?
If you observe coolant leaking between the handpiece and connectors of your iM3 Media Converter, it's probably due to a defective O-ring in the handpiece connector. You should replace the O-ring.
Brand | iM3 |
---|---|
Model | 42-12 |
Category | Media Converter |
Language | English |
Turning the unit on/off and activating the handpiece with coolant flow.
How tip vibration is activated and power settings are adjusted.
Emphasizes using the tip parallel to the tooth surface to avoid enamel damage.
Warnings about contact with restorations, mucosa, and the benefits of ultrasonic sound.
Instructions not to sharpen, bend, or reshape the scaler tips.
Diagnosing and resolving coolant leaks, spray, vibration, and power problems.
Guidelines for cleaning tip inserts, handpiece, and control box.
Key technical data including connection, voltage, fuses, and consumption.