6 - 9
PN 074-519-P1C
3000 Micro GC Operating Manual
6.4 Change Carrier Gas Configuration
The type of carrier gas used for analysis can be changed through the embedded
web page.
NOTE: If a carrier gas type is changed without following this procedure, inverted
peaks will appear in the chromatogram.
1 Open the web browser.
2 In the address bar, type the current IP Address of the instrument. Press Enter.
See Figure 6-9.
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
Figure 6-9 Internet Explorer address bar
3 Click the Gas Type tab. A window will display that is similar to Figure 6-10. The
current gas configuration for installed GC modules is shown.
Figure 6-10 Example gas type tab