6 - 11
PN 074-519-P1C
3000 Micro GC Operating Manual
8 Select the new carrier gas type from the drop-down list. See Figure 6-13.
Figure 6-13 Selecting a new carrier gas type
9 Click Submit. This turns the column heaters off (to cool the columns and avoid
thermal shock) and displays the Check Status view. See Figure 6-14.
Figure 6-14 Check status view
10 Click Check Status to display a window which shows the current column
temperature(s). When the column temperatures are below 60
ºC, Close the
Check Status window. See Figure 6-15.
NOTE: Changes may be canceled up to this point by closing the browser or by
using the browser back feature.
Figure 6-15 Example check status window
Periodically refresh display
Actual column tempertures