8 - 26
PN 074-519-P1C
3000 Micro GC Operating Manual
8.3.4 Running a High Pressure Gas Sample without Liquids/Particulates
1 If the Pressure Reducer (PN G2815A) is not installed, install one as instructed
in section 5.1.4, Installing the Pressure Reducer, on page 5-8. See Figure 8-31.
NOTE: The Pressure Reducer can connect directly to the 3000 Micro GC or to
the sample inlet filter (PN G2801-60900).
Figure 8-31 Pressure reducer sample conditioner (PN G2815A)
2 Turn 3000 Micro GC on.
3 Check that the carrier gas supply is sufficient to run all of the samples.
4 If using a sample vessel, prepare the sample.
5 Connect the sample conditioner or gas source to the sample input fitting on the
Pressure Reducer accessory.
6 Open the sample vessel valve.
7 Use EZ IQ software to load the desired method and perform the analysis.
8 When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.
9 Repeat steps 4 to 8 to run the next sample.
Sample in
To sample input fitting
on 3000 Micro GC
Vent
Adjustment knob
Mounting bracket not shown