Inficon 3000 Micro GC - RGA Checkout Gas

Inficon 3000 Micro GC
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 - 19
PN 074-519-P1C
3000 Micro GC Operating Manual
11.5.1 RGA Checkout Gas
Refinery Gas Analyzer (RGA) Checkout Gas may be used during installation of a
3000 Micro GC configured for refinery gas composition analysis to verify chemical
performance. Table 11-11 lists the component concentrations.
NOTE: Do not use RGA Checkout Gas as a calibration standard for specific
refinery gas applications. Since refinery gas composition varies greatly
among various petrochemical processes, acquire a calibration standard
that meets specific requirements.
Table 11-11 RGA checkout gas
Components Concentration (Mole %)
hydrogen 12.0%
oxygen 0.2%
nitrogen balance
carbon monoxide 1.0%
carbon dioxide 3.0%
methane 5.0%
ethane 4.0%
ethylene 2.0%
acetylene 1.0%
propane 2.0%
propylene 1.0%
1,2-propadiene 1.0%
methyl acetylene 1.00%
isobutane 0.3%
n-butane 0.3%
1-butene 0.3%
isobutylene 0.3%
trans-2-butene 0.3%
cis-2-butene 0.3%
1,3 butadiene 0.3%
isopentane 0.1%
n-pentane 0.1%
1-pentene 0.1%
cis-2-pentene 0.1%
trans-2-pentene 0.1%
2-methyl-2-butene 0.05%
n-hexane 0.05%