5 - 7
PN 074-519-P1C
3000 Micro GC Operating Manual
11 Install one end of the sample inlet tubing, 1/4 in. nut, and front and back ferrules
provided in the kit onto the sample input fitting on the bottom of the filter as
shown in Figure 5-6. Connect the open end of the sample inlet tube to the
sample.
Figure 5-6 Installing the sample tube
WARNING
If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.
12 Uncoil the vent tubing. Run the exhaust to a fume hood or other designated
area for hazardous waste disposal.
Figure 5-7 shows the Genie Filter Assembly mounted on 3000 Micro GC.
Figure 5-7 Genie filter assembly mounted on 3000 Micro GC
Sample outlet tubing
Sample inlet tubing
Vent tubing
Genie filter assembly
G2817A on a 2-channel 3000 Micro GC shown