8-7 Maintenance
kina22e chapter 8.fm technical handbook(1206)
6 Insert new operation fluid reservoir into the pump up to the O-Ring (see Fig. 8-6).
7 Re-assemble bottom lid. Make sure that the O-ring remains in the correct position,
as otherwise large leaks can occur.
Notice: The bottom lid will align the operation fluid reservoir to the correct axial
position.
8 From the main menu go to HISTORY & MAINTENANCE / CONFIRM MAINTENANCE and
confirm with the appropriate button on the right side of the display that the operation
fluid reservoir has been replaced / cleaned.
The actual date and time as well as the actual number of operating hours will be
saved and a reminder for maintenance will be issued again after 3 years.
Notice: The Ecotec E3000 must be set to
ADVANCED mode for confirming
maintenance tasks.
Fig. 8-6 O-Ring of operation fluid reservoir
Do not completely push in the operation fluid reservoir. The operation fluid reservoir
is correctly positioned by the cover after screwing it in.