Maintenance 8-10
kina22e chapter 8.fm technical handbook(1206)
3 Push in new filter pads.
Notice: The metal grid is not included with the spare filter pads. Please clean the
metal grid carefully and use it again.
4 Switch on the Ecotec E3000.
5 Hold finger against the capillary filter:
You should be able to feel the flow.
6 If warning “Flow through capillary too low” remains, replace the sinter filter (see
Section 8.3)
7 From the main menu go to HISTORY & MAINTENANCE / CONFIRM MAINTENANCE /
S
NIFFER TIP FILTER and confirm with the ok button on the right side of the display that
the sniffer tip filter has been replaced. The actual date and time as well as the actual
number of operating hours will be saved and a reminder for maintenance will be
issued again after the preset number of hours. For details see Chapter 4.7.
Notice: The Ecotec E3000 must be set to
ADVANCED mode for confirming
maintenance tasks.
Fig. 8-9 Pushing out filter pads
Fig. 8-10 old and new filter pads

Table of Contents