General Information 1-15
kina22e chapter 1.fm technical handbook(1206)
1.6.2 Mechanical Connections
ECO-Check
Reference leak (optional)
Please insert the ECO-Check reference leak into the opening in the housing of the
main unit. Make sure that the Sub-D plug is properly connected with the ECO-Check
leak.
Notice: When properly inserted, the ECO-Check reference leak will still protrude by
approx. 10 mm.
On first usage of your ECO-Check you need to initialize the use of this reference leak
in the Ecotec E3000 software menu.
Please perform the following steps:
1 Insert the ECO-Check into the appropriate opening of the Ecotec E3000
2 In the software menu go to HISTORY & MAINTENANCE / REPLACE ECO-CHECK.
Notice: The Ecotec E3000 must be set to
ADVANCED MODE for access to this
submenu (See section 4.5.3). Go to
SETTINGS / DISPLAY / USER MODE)
3 On the certificate, which is delivered with the ECO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the second line and press OK.
Notice: The ECO-Check reference leak must be installed in the Ecotec E3000
when pressing
OK.
SL3000 Sniffer line
In order to operate the Ecotec E3000 it is essential for the sniffer line to be
connected. The connection for the sniffer line is located at the front of the Ecotec
E3000 left of the ECO-Check reference leak.
Insert the plug into the opening with the red dot on the plug and the slot in the front
cover aligned until the connector engages. To disconnect the plug, retract the
coupling and remove the probe’s line.
Fig. 1-4 Initializing the ECO-Check reference leak