How the Instrument Works 2-7
kina22e chapter 2.fm technical handbook(1206)
In cases where the main unit is placed in a difficult or inconvenient to access area
the built-in ECO-Check reference leak can be removed from the main unit for easier
access and connected to the main unit via the Sub-D connector with a commonly
available extension cord. The ECO-Check reference leak can then be placed in an
area where it is convenient for the operator to be reached.
Notice: The ECO-Check reference leak is not part of the Ecotec E3000 shipment
and needs to be ordered as a separate part no. (see section 1.3.3).
Notice: If you have not purchased the ECO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to
SETTINGS / INTERFACES /ECO-CHECK and set the ECO-Check to “DISABLED
to prevent future warnings (see section 4.6.7)
The Ecotec E3000 must be set to
USER MODEADVANCED“ for access to this
submenu (See section 4.5.3). Go to
SETTINGS / DISPLAY / USER MODE.
Fig. 2-8 Connections for remote use
Pos. Description Pos. Description
1 Connector at built-in test leak 2 Connector at
Ecotec E3000 housing