3-12 Operation of the Ecotec E3000
kina22e chapter 3.fm technical handbook(1206)
The sniffer line display will switch to red background color if the global trigger is
exceeded (or at least one if two gases are selected for the program used).
The gas exceeding the global trigger is displayed in the results list. However, you can
switch between the results of the two different gases with the "A" and "B" buttons on
the right side of the display.
The display of the sniffer line will show the results for the two different gases
alternating automatically.
By pressing the right probe button the next testing cycle may be started.
3.4.3 The Info Page
By pressing the INFO button on the main display a general info page will open.
Information on the software version used, the number of operating hours, the serial
number, time & date and the currently selected alarm type will be stated. Also, the
selected gases with their mass position will be listed. Gases that are disabled will be
marked as such. For gases that are enabled also the currently selected trigger level
will be started.
If an active warning exists this will be displayed in the info page instead of the gas
info lines as shown below.
Fig. 3-14 Result of I•Guide program: unit under test failed
Fig. 3-15 Info page without errors or warnings