Important Safety Precautions 0-7
kina22eIVZ.fm Technical Handbook(1206)
Failure to observe the following precautions could result in serious personal
injury:
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Ecotec E3000 with the ground conductor unconnected is not
permissible.
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
close distance as this may cause permanent damage to the eye.
Danger of electric shock.
Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not
be exceeded or reached to ground or ground equipment conductors.
According to the type of in- or outputs lower voltages had to be accepted. For this,
please refer to the information given in the responding chapters.
For all maintenance on the Ecotec E3000, the Ecotec E3000 must be disconnected
from power.
Before exchanging the air filter the Ecotec E3000 must be disconnected from
power.
Before exchanging the fuses the Ecotec E3000 must be disconnected from power.