4-30 Equipment Settings
kina22e chapter 4.fm technical handbook(1206)
4.7 The Info Menu
The INFO MENU lists all internal data that may be useful for any trouble-shooting of the
Ecotec E3000. The info menu consists of 8 pages. On each page the upper left
button “
BACK” allows to leave the info menu, the lower left button allows to go back
one page whereas the lower right button allows to go to the next page. The page
number is displayed in the upper right corner.
Menu item Format Description
Page 1: General Data
Foreline pressure mbar Pressure at inlet to mass
spectrometer
Flow sccm Flow through sniffer line
Total pressure mbar Pressure in mass
spectrometer
Time since power on Min
Operation hours h
Serial number 9000 XXX XXXX
Software version x.xx.xx
Circuit temperature °C / °F Temperature of main board
TSP temperature °C / °F Temperature of mass
spectrometer
Test leak temperature °C / °F Temperature of ECO-Check
reference leak
Page 2: Turbo molecular pump data
State On/off
Current error code
Rotation speed Hz
Current A
Voltage V
Power W
Operation hours TMP H Operation hours of TMP
Operation hours TC H Operation hours of frequency
converter
for TMP